Olis Closed-Cycle Cooling System Problems

Symptom: Lamp keeps turning off

Probable Cause: The flow volume in the closed-cycle cooling system is too low.

SOLUTION: Adjust flow volume.

Leave the lamp turned off and leave the CCCS hooked up and plugged in.

Open the CCCS lid by turning the screw-top latch with a flat-head screwdriver.

Lift lid and locate the valve with the clear plastic handle.

Turn the valve CLOCKWISE to shut off the flow completely (the red light on the front of the CCCS should go out, if it is not already out). This establishes a starting point for flow adjustment.

Turn the valve COUNTER CLOCKWISE just until the red light on the front of the CCCS comes on. Then, turn the valve COUNTER CLOCKWISE about 1/16 of a turn further (just enough to ensure that the red light will stay on).

Close the lid and relock the hatch.

Read more about the Olis Closed Cycle Cooling Box.

 