Paul's Practical Guide - page 7

Troubleshooting Guide

Symptom Possible Problem Solution
Lamp difficult to light Old lamp(1,000 hours) Change lamp
Current control on power supply Increase power supply current
Lamp shuts off repeatedly Bubble in cooling box fluid Use "burper bottle" to remove air from system
PEM Error Hinds PEM unit is not powered or connected to C4 on A/D board (computer) Turn Hinds unit on and ensure connections are secure
PEM HV increases to Maximum (1100 V) PMT detectors seeing no light Ensure the lamp is on
Slits are open
No inappropriate filters are in place
The sample is not opaque at the current wavelength
Very high PMT HV in far UV UV-absorbing buffer components Remove UV-absorbing components
Old lamp Replace lamp
Too much oxygen in system Increase nitrogen purging
Irreproducible signal intensities Sample sticking to cuvette Clean cuvette
Increase concentration
Change buffer components to avoid sticking
Sample degraded by measurement beam Reduce slit width
Reduce measurement time
Increase sample concentration
CD intensity not correct (or upside down) Phase adjust incorrect Change phase adjust value to give maximum signal
PMT cables reversed Plug in cables correctly

 