Sample Data from the DM-45 Spectrofluorimeter
Comments:
Collection Mode: Scan
Number of Points: 200
Excitation Monochromator = Fixed @ 360 nm
Emission Monochromator = 370 to 470 nm
Timing Mode: Constant Time
Integration Time: 1.0 sec.
Reduction Mode: Photon Counting
Scan Mode: Fix Everything (Photon Counter)
Total Elapsed Time = 6:40 (min:sec)
Anthracene 0.6 mm all around
Click image to enlarge in new window. |
 |
Comments:
Fit Model: A -> B
Number of Species: 2
Kinetic Std Dev: 1.59E-02
Overall Std Dev: 7.52E-03
Rate Constant 1: (4.8±0.8)*E+01
Iterations: 1
Click image to enlarge in new window. |
 |
Comments:
Collection Mode: Scan
Number of Points: 320
Excitation Monochromator = Fixed @ 342 nm
Emission Monochromator = 440 to 600 nm
Timing Mode: Constant Time
Integration Time: 1.0 sec.
Reduction Mode: Photon Counting
Scan Mode: Fix Everything (Photon Counter)
otal Elapsed Time = 5:24 (min:sec)
Ovalene standard 2 nm slits
Click image to enlarge in new window. |
 |
Comments:
Collection Mode: Scan
Number of Points: 200
Excitation Monochromator = Fixed @ 550 nm
Emission Monochromator = 555 to 655 nm
Timing Mode: Constant Time
Integration Time: 1.0 sec.
Reduction Mode: Photon Counting
Scan Mode: Fix Everything (Photon Counter)
Total Elapsed Time = 3:23 (min:sec)
Rhodamine Standard 2 nm bandpass
Click image to enlarge in new window. |
 |
Comments:
Collection Mode: Scan
Number of Points: 400
Excitation Monochromator = Fixed @ 350 nm
Emission Monochromator = 370 to 470 nm
Timing Mode: Constant Time
Integration Time: 1.0 sec.
Reduction Mode: Photon Counting
Scan Mode: Fix Everything (Photon Counter)
Total Elapsed Time = 6:39 (min:sec)
Raman 1.24 mm slits
Click image to enlarge in new window. |  |
Comments:
Collection Mode: Scan
Number of Points: 120
Excitation Monochromator = Fixed @ 350 nm
Emission Monochromator = 420 to 360 nm
Timing Mode: Constant Time
Integration Time: 1.0 sec.
Reduction Mode: Signal
RC Filter: 5 msec
Repeat Scans as a fxn of: None (5 scans; 1 shown)
Scan Mode: Fix Everything (PMT 1 HV=1100)
Total Elapsed Time = 10:29 (min:sec)
Raman of water 1 cm cell, 1.24 mm slits, **PMT detector**
Place your cursor over the image to see the 3D data representation. Click image to enlarge in new window. |
 |