Wavelength Range
|
190 to 820 nm |
|
Photometric Noise
|
< ± 0.05 nm, NIST 2034, ten consecutive measurements
|
|
Wavelength Accuracy
|
< ± 2 nm
|
|
Slit Width
|
2 nm
|
|
Stray Light
|
at 200 nm; < 0.05%
at 340 nm; < 0.05%
|
|
Photometric Accuracy
|
±0.005 Å; at 1 Å, 440 nm, NIST 930e
|
|
Photometric Noise
|
<0.0002 Å; at 0 Å, 500 nm, sixty 0.5-second scans, RMS
|
|
Photometric Stability
|
<0.002 Å/hr; (at 0 Å, constant ambient temperature, after 1-hour warm up): measured over 1 hour, every 5 seconds
|
|
Baseline Flatness
|
<0.001 Å; 0.5 second blank, 0.5 second scan, RMS
|
|
Shortest Scan Time Rapid Scan Mode
|
0.1 seconds; full range
10 scans per second; 70 nm wavelength range
NEW! Added by Olis® computer control!
|
Power Requirements:
Line Voltage
Line Frequency
|
80-132 or 170-264 V AC
47-66 Hz
|
|
Dimensions
|
Length: 23.8 inches
Width: 16.7 inches
Height: 7.3 inches
|
|
Weight
|
33.2 pounds
|
Environmental Conditions:
Operating Temp.
Non-operating Temp.
|
0-40 °C (32-104 °F)