The PE 983 Data Collection Scheme Is Shown in Figure 1
The phase relationship between PSC and BSC in the PE983 produces the sequence of signals shown:
Figure 1

This data collection format causes temporal separation between acquisition of sample and reference signals.
The Olis 983 Data Collection Scheme Is Shown in Figure 2
Figure 2

This sequence results in minimal temporal dispersion between sample and reference signals. Now the signals are R2, S1, DR, and S2, R1, DSthat is, the sample and reference signalsare as close together in time as possible, minimizing the effects of a change in the detector's output. Only the order of the six numbers is altered. The calculation (and thus the answer) is the same, but with a reduced potential for error.
The six values collected as just outlined, are used to calculate %T:
Equation 1

PREVIOUS | 1 2 3 4 | NEXT
|